Different results with glucose measurements?

We regularly receive questions about a difference in glucose measurements. A glucose meter can give a different result at 10 o'clock in the morning than at 6 o'clock in the evening. Two glucose meters next to each other will also have an error. In principle this is no problem at all. Even if you were to measure the same drop of blood on two meters at the same time, you might find an deviation. The meters are then not suddenly unreliable. But what exactly is this?

Deviation in glucose meters

Every glucose meter has a deviation (spread), with most meters this is a maximum of 15 percent of a reference value. This 15 percent is described in the glucose meter consensus document, which serves as a quality reference in the Netherlands. This describes what requirements a glucose meter must meet in order to be reimbursed by insurers (ISO standard). All meters that you find on diabetesmagazijn.nl comply with this ISO standard. They therefore never deviate more than 15 percent from the reference value.

For low glucose values ​​(glucose < 5,55 mmol/L), the deviation is calculated in absolute numbers: the absolute deviation may not be greater than 0,83 mmol/L.

As the results of two blood glucose meters are compared with each other, the above standards do not apply, because you do not know what the reference value is.

Compare glucose meters

If you want to compare glucose measurements, you should only do so with laboratory tests. If you compare two glucose meters with each other, they can both differ and you still do not know which one differs and by how much. Optimal is when compared with a well-defined reference method. In practice, laboratory research suffices.

Suppose the reference value (a value measured in the laboratory, in the same sample, at the same time) turns out to have a value of 10 mmol, then a glucose meter may indicate a value between 8,5 mmol and 11,5 mmol and then it is still correct.

The meter meets the ISO standard with the above calculation example and is therefore reliable. Before treatment, it is of the utmost importance to always check the insulin setting of people with diabetes with the same meter. And the advice is also to look at averages or repeat measurements, so not to immediately draw conclusions based on one measurement.

Differences in measurements occur. But this is almost never due to the meter. The causes can be sought, for example, in different moments of measurement or strips that have been exposed to moisture or light. If you are in doubt, you can try to take measurements with control solution, but it is best to compare the measurement with laboratory research.

6 thoughts on “Different results with glucose measurements?”

  1. Dear Elke, thank you for the quick response, I understand the assumption that there is a difference in the various meters. The difference between my meter and that of the GP is just as great as that during the lab measurement.
    Lab measurement 9.1, 15 minutes later my meter 11,2, at the doctor's 9,2, same time for me 12,2
    I took the trouble to make this comparison because I would like to know how my health is going. If it turns out that the meter is not good, it has no value.

    I would still like a response M. Brink

    1. diabetesmagazijn.nl

      Dear MGM Brink,

      I think this response is a response to an email, right?
      Would you like someone from Diatesse to contact you by telephone?

      Could you please email your answer to info@diabetesmagazijn.nl?

    2. I would like to know between the glucose measurements. Nordafrica values ​​and Europe.

  2. Nevertheless, measurements were carried out again. Now on a Ht one the measurement was 8.1. Measured twice on the Prime Minister, the measurement was 8.0 and 8.2.
    Seems fine to me now. Perhaps the temperature of the hands played a role.
    Thanks for your response.
    Greetings: Piet van Nief

    1. Good evening. Ok, the Freestyle meter measures this, it shows lo, while a battery check meter shows 12.2. What is correct?

  3. G. van der Kooi

    Good day,
    I regularly check my blood glucose levels with Accuchek performa. Always fasting between 6,2 and 6,8. Now suddenly at the doctor 8,2. She indicates a 15% spread but still... Who is right? Annual lab test gives 6 without decimal place. Still okay?